Deciding whether to purchase either a viscometer or a rheometer is not always straightforward. This guide will take you through the differences and how a rheometer, such as our Rosand capillary rheometer range, could be more suited for your needs.

What’s the difference?

Typically a viscometer employs a mechanical bearing that limits the speed and torque capabilities of the instrument, whereas a rheometer uses a low friction air bearing. This means a viscometer can be a solution for material, process or production tests that require simple flow measurements on Newtonian materials (where viscosity is independent of shear rate), however the performance of a rheometer allows far greater characterization of flow, deformation and even the tackiness of a material (for Newtonian and non-Newtonian materials). 

A viscometer can offer portability for field or remote testing. Rheometers, while generally more expensive than viscometers, are more versatile and have a much wider dynamic range of control and measurement parameters.  

rosand-rheometer.jpg

Measuring viscosity

Measuring viscosity is the most common application required for a viscometer or rheometer. For most products, the viscosity is required to be high at low shear rates to prevent sedimentation or slumping, but to thin down at higher shear rates to facilitate application or processing. Hence a single viscosity measurement is not sufficient to describe the viscosity of such materials and the viscosity should be measured over a range of shear rates or stresses 

Typically, a viscometer can measure in the range of about 0.1 to 103 s-1 while a rheometer extends the measurement range from 10-6 to 105 s-1. A broader measurement range enables relevant data to be obtained by exposing the sample to conditions that are realistic to the conditions applied during product manufacture or use. 

Processes such as sedimentation are best suited to analysis with a rheometer due to its low torque capabilities. High speed control of a rheometer also allows analysis of a very high shear rate process, such as spraying.

ProcessMinimum shear rate (S-1)Maximum shear rate (S-1)ViscometerRheometer
Reverse gravure105106✔ 
Spraying104105✔ 
Blade coat103105✔ ✔ 
Mixing/stirring10103✔ ✔ 
Brushing10103✔ ✔ 
Pumping1103✔ ✔ 
Extrusion1102✔ ✔ 
Curtain coating1102✔ ✔ 
levelling10-20.1✔ 
Sagging10-20.1✔ 
Sedimentation10-610-2✔ 

Yield stress 

After viscosity, yield stress is probably the more routinely measured rheological property because many consumer products gain value from having one.  

Yield stress varies as a function of the temperature and timescale during which the stress is applied. Rheometers can provide more relevant yield stress data than a viscometer by enabling application of the broad range of these methods. Applying a stress ramp is looked upon as the easiest way with a rheometer. 

Which is best for me? 

The increased versatility and performance make rheometers an excellent tool for research, product and process development, as well as quality control testing. Both viscometers and rheometers are complementary, and it is not uncommon within a single organisation to find viscometers used for QC testing on products that have been developed using a rheometer. 

Rosand Linie

Rosand Linie

Kapillarrheometer zur prozessrelevanten Materialprüfung.

Mehr Details
Typ der Messung
Rheology and viscoelasticity
Technologie
Rheometrie - Kapillar
Temperaturbereich 5°C - 500°C