The Zetasizer AT combines Malvern Panalytical's industry leading Dynamic Light Scattering (DLS) technology with two decades of expertise in design and automation of process interfaces for analytical instrumentation.

The Zetasizer AT is a fully-automated nanoparticle characterization system, delivering lab quality results on-line and in a timely manner for process optimization.

  • Ability to monitor processes involving nanoscale particles or droplets in real time e.g. homogenization
  • Ability to track changes in particle size with time and determine when the end point is reached
  • Automated sampling system facilitates sample extraction, degassing and dilution to the required concentration minimizing operator input
  • Sampling system designed to minimize sample consumption
  • Easy transfer of laboratory DLS methods to on-line system
  • Utilizes Malvern's Universal Interface for system integrators (Malvern Link II)

An automated sample preparation system utilizes high precision pumps and flow control valves to extract a sample out of a process pipe or vessel and dilute it to the required concentration. The sample is then degassed and pumped through the analyzer’s flow cell. After analysis the sample is discharged and the cycle repeats. Results are immediately available on the screen of the system computer and can be linked to plant SCADA or LIMS system.


Malvern Link II software controls every aspect of the sampling, dilution and degassing steps using Standard Operating Procedures (SOPs).

The Zetasizer AT also allows easy batch measurements of samples from other areas in the process, effectively doubling as a laboratory system.

Measurement range:
1nm – 1 micron* (diameter) (0.3nm – 10 microns*  at-line)
Measurement principle:
Dynamic Light Scattering
Accuracy:
Better than +/-2% on NIST traceable latex standards
Precision / Repeatability:
Better than +/-2% on NIST traceable latex standards
Sensitivity:
0.1mg/mL (Lysozyme)
Sample Interface:
1/16" OD (0.040" ID) or 1/8" OD (1/16" ID) tube
Pressure:
0 - 3 Bar (0- 44 psi) at sample inlet.
Temperature:
0 – 90 °C (flow cell only)
Minimum sample volume:
4ml (application dependent)
Dilution:
up to 1:50 / infinite (application dependent)
Power:
110-240VAC, 6A max, 50-60Hz
Compressed Air for optics purge (optional):
2-8 Bar (29-116 psi) per ISO8573-1:2010 [1.7.2]
Ingress Protection (IP) rating:
IP54
Dimensions (W, D, H):
98 x 66 x 95 cm (38.6 x 26 x 37.4 inch)
Weight:
120.5 kg (266 lb)
Conforms to:
Low Voltage Directive 2006/95/EC
EMC Directive 2004/108/EC
Pressure Equipment Directive 97/23/EC
BS EN 61010-1:2010     BS EN 60204-1:2006 + A1:2009   BS EN 61326-1:2006 
CE marked

Protect your investment and ensure optimized performance at all times with Malvern Panalytical's service plans.

When you purchase a Malvern Panalytical product we understand that this is just the first stage of a working relationship that will last for the lifetime of the instrument. Depending on your needs, Malvern Panalytical will provide the support for your business.

Help Desk

Training

Software downloads

Malvern Panalytical's service options- choose the right plan for you:

Malvern Panalytical Platinum Plan

For laboratories where maximizing instrument up-time is critical to its daily performance. Our highest level of response and the 'all inclusive' price plan takes care of those unexpected repair bills should the unfortunate occur.

Malvern Panalytical Gold Plan

Enhance your laboratory productivity by maintaining instrument efficiency. Still receive that priority response, combined with specialist technical and software support, we'll keep your instrument performance on track to deliver.

Malvern Panalytical Bronze Visit

For laboratories looking to optimize instrument performance, regular preventative maintenance is essential.

Platinum PlanGold PlanBronze Visit
Annual PM/PV Test
yes
yes
no
Telephone / Email Support
yes
yes
no
Priority Response
yes
yes
no
Emergency Breakdown visits*
yes
yes
no
Parts Included
yes
no
no
Technical & Software Support
yes
yes
yes
IQ/OQ (Pharma)***
yes
yes
yes

*including labour & travel costs***available at an additional cost