Zetasizer Red Sensitivity – How much protein do I need for the Zetasizer red?

Zetasizer red sensitivity calculator, Minimum concentration calculator for any protein in the Zetasizer Advance Red Label

You have been asking for a while, so here it is! The Zetasizer Red sensitivity calculator! It predicts the minimum concentration in order to measure your protein. There are different values for the size from backscattering, side scattering, or multi-angle MADLS scattering. And as a bonus, the sensitivity calculator also estimates the hydrodynamic size of a typical globular protein. All you have to do is enter the molecular weight.

Why a sensitivity calculator for light scattering?

The intensity of light scattering depends strongly on the size of a molecule or particle: the larger the object, the more light it scatters. In order to acquire good statistics, we then want to ensure enough light arrives at the detector. This is related to the size of the object. We can estimate the size of a protein molecule from its molar mass or molecular weight. In order to get to this estimate, we assume a typical globular shape. The calculator gives you both the size and the minimum concentration for a Zetasizer Advance (Red Label) system, specifically for the Zetasizer Red Ultra/Pro/Lab models.

How do I use the calculator for the Zetasizer red?

The first input box in the sensitivity calculator is for the molecular weight of your protein. Simply enter the expected molar mass of your molecule in kilo Dalton [kDa]. However, if you only know the number of amino acids, you can estimate a typical amino acid is about 0.11kDa. Similarly, if you have the number of base pairs, a typical is about 0.65 kDa. Once you enter a number in the molecular weight box, the estimated radius and sensitivity values automatically update in your browser.

What are the different sensitivity values?

There are basically two kinds of sensitivity:

Hope this gets you to the correct minimum concentration for your protein! Thanks for using the Zetasizer red sensitivity and size calculator for proteins.

Protein Sensitivity Calculator (Red Label)

This calculator shows the minimum required protein concentration for the Zetasizer Advance Red Label Series.

Please enter the molecular weight in kiloDalton

This is the molar mass in kDa = 1000 Dalton. For estimates: 1 amino acid ≈ 0.11 kDa, 1 base pair ≈ 0.65 kDa.

Calculated radius and concentrations

The result is the minimal concentration of a protein, required to measure in the Zetasizer Advance Red Label Lab/Pro/Ultra. This calculator estimates the radius of the protein, and predicts the minimum required concentration, for various Zetasizer models, as indicated.

Note: this is the radius, not the diameter.

All concentration values are in mg/mL (micrograms per microliter or grams per liter).

Size estimate from molar mass is based on empirical relationships:

U. Nobbmann, “Characterization down to Nanometers: Light Scattering from Proteins and Micelles”, in Mesoscale Phenomena in Fluid Systems, ed. F. Case and P. Alexandritis. ACS Symposium Series 861, pp. 44–59 (2003).

Thanks for using the proteins sensitivity and size calculator for the Zetasizer Advance series (Red Label).

Previously

Have any questions? Please email me ulf.nobbmann@malvern.com – Thanks! Opinions are those of the author. Our editorial team modifies them occasionally

Este artigo pode ter sido traduzido automaticamente


Thanks for using the proteins sensitivity and size calculator for the Zetasizer Advance series (Red Label).

Previously

Have any questions? Please email me ulf.nobbmann@malvern.com – Thanks! Opinions are those of the author. Our editorial team modifies them occasionally

" }