PIXcel3D

The first detector to bring 0D-1D-2D and 3D data to your diffractometer

The PIXcel3D is a unique 2D solid-state hybrid pixel detector. Each pixel is 55 microns x 55 microns and the detector array is 256 x 256 pixels. The detector, now based on Medipix3 technology, brings unrivalled signal to noise with its point spread function of one pixel and multiple energy discrimination levels.

Solicite um orçamento Solicite uma demonstração Fale com Vendas

Overview

The first detector to bring 0D-1D-2D and 3D data to your diffractometer

The PIXcel3D is a unique 2D solid-state hybrid pixel X-ray detector. Each pixel is 55 microns x 55 microns and the detector array is 256 x 256 pixels. The detector, now based on Medipix3 technology, brings unrivalled signal to noise with its point spread function of one pixel and multiple energy discrimination levels.

Medipix3 technology brings:

  • An increase in range of detection energies from Cr to Cu
  • Improved energy resolution
  • Increased linearity range
  • Improved detection efficiency
  • Documented (published) radiation hardness studies.

Specification

TypePIXcel3D Detector with Medipix3
Window size14 mm x 14 mm
Efficiency Cu Kα>95 %
99% Linearity range0 – 6.5 x 109 cps – Overall
0 - 25 x 106 cps - Column
Energy resolution around Cu Kα18%
Maximum count rate30 x 109 cps – Overall
120 x 106 cps - Column
Maximum background<0.5 cps - Overall
Active length14 mm
Smallest step size0.0016º 2θθ at 240 mm goniometer radius
Supported wavelengthsCu, Co, Fe, Mn, Cr
Point Spread Function (PSF)1 pixel (FWHM)
Defective pixels< 1%