Conductivity – check 3 points for better zeta

In electrophoretic light scattering measurements we apply a voltage and measure a current. This current depends on the conductivity of the sample. But what is this conductivity? Is there a good conductivity we should target or a bad conductivity to avoid? How many milli Siemens per centimeter or mS/cm are good for the ZTS1240 transfer standard?

What is conductivity?

Conductivity is the ability of a liquid to conduct electricity, or in other words to allow an electric current to flow through the liquid. In our instrument, we apply a voltage V and measure a current I and the resulting conductivity G is G = I/V, with units Siemens. From physics, you may be more familiar with the resistance R = V/I with units Ohm. So resistance is the inverse of resistance. For more general details take a look at Conductivity (electrolytic) in Wikipedia.

In practical terms, conductivity tells us how much salt, or how many ions are in a liquid. The higher the conductivity, the more electric current can flow through the sample, that means more ions are present in this sample.

Suitable range for conductivity?

Completely de-ionized water has a conductivity of 0.055 μS/cm (This is the inverse of 18 MegaOhm⋅cm, the well-known ultra-clean milliQ water limit). The Zetasizer software reports conductivity in units of milli Siemens per cm, so on that scale, deionized water has a conductivity of 0.000 mS/cm: we cannot detect any ions in the system. For reasonable zeta measurements, it is advised to have some ions present in the sample a concentration of for example around  0.001 M NaCl is normally considered adequate. For very low conductivity, polarization/field distortion effects can lead to non-ideal results.

What is the conductivity for ZTS1240 standard?

The ZTS1240 Zeta Transfer Standard is a formulated preparation of latex particles in a pH 9.2 buffer and contains some ions. All buffers contain ions to establish their buffering capability. The typical range of conductivity in the ZTS1240 transfer standard is 0.31 ± 0.039. So we consider the conductivity check to pass if it falls in between 0.271 mS/cm and 0.349 mS/cm as the lower and upper limit.

Conductivity pass fail criteria for the ZTS1240 Zeta transfer standard from Malvern for the Zetasizer

Further relevant posts involving conductivity

Este artículo puede haber sido traducido automáticamente.