Regular maintenance by the user of MicroCal Auto-iTC200 is essential for performing high quality experiments and for obtaining high quality results. This note offers advice on a routine maintenance schedule.
Regular maintenance by the user of MicroCal Auto-iTC200 is essential for performing high quality experiments and for obtaining high quality results. Cell cleanliness has a significant impact on data quality.
| Operation | Necessary reagents, parts, and tools |
|---|---|
| Cell cleaning | 20% Contrad 70 or 14% Decon 90 |
| Filling the reference cell | Sample loading syringe, 28428942 (SYN50018P) |
| Replacing the plunger tip | Soft-grip tweezers, 28428966 (TOL50165P) Tip removal tool, 29012502 (TOL021706) Tip insertion tool, 28428964 (TOL25003) Tip, 100ul plunger (pk/5), 28426097 (TIP130010-005) |
| Cleaning the titration syringe | Soft-grip tweezers, 28428966 (TOL50165P) Wire syringe cleaning, 28429013 (WIR020598) Brush .060 IN, 28428967 (TOL501660) Brush .200 IN, 28428968 (TOL501670) |
| System verification | EDTA test kit, Auto-iTC200, 28429064 (KIT020989) |
Malvern Instruments provides Preventive Maintenance service and will be happy to discuss the maintenance interval most suitable to your situation. Contact your local representative or find your local contact details here.