NanoSight LM12 Power Supply Safety Notice

This guide will run through checks to carry out on your NanoSight LM10 power supply, maintenance required, and replacement details.

This guide will run through checks to carry out on your NanoSight LM10 power supply, maintenance required, and replacement details

The laser module (LM12) is classified (to BS EN 60825-1) as a Class 1 laser device. However, this illumination unit contains a Class 3B laser source. In order to be able to classify the laser module as a Class 1 laser product, an interlock tool has been incorporated in the form of a security hex driver bonded to the cable.

If the cable is beginning to break down you could see an intermittent loss of laser power on your NanoSight LM10 instrument.

로그인

Not registered yet? Create an account