Oil-Trace

準確的微量元素分析

Oil-Trace 是一個由一套標準組成的軟體包,使用戶能夠使用 WDXRF 和 EDXRF 對液體進行準確可靠的元素分析。 Oil-Trace 解決了分析液體時的兩個問題:

  1. 在存在未知 CHON 元素混合的情況下進行充分的矩陣校正
  2. 修正因分析樣品體積變化而產生的錯誤

透過有效解決這些問題,Oil-Trace 為石油分析帶來了顯著的好處。

尋找更多資訊?

如需報價、更多資訊或下載手冊,請選擇下面的選項。


Features

Corrections for a mix of light elements

The matrix of a liquid can contain compounds that cannot be not measured by XRF (e.g. -CH2-, -CHOOCH2- or -CH2OCH2-). The matrix effects of these compounds, however, are very different and influence the accuracy.

For an accurate analysis of elements like sulfur the determination of these compounds is crucial. These compounds have a high influence on the Compton scattered tube intensity. For this reason the FP model is extended with a model to calculate the theoretical intensity of scattered tube intensity.

Oil-Trace exploits this advanced FP model to correct for the influence of these compounds, resulting in increased accuracy for elements like Mg, Al, Si, S and P. 

Correcting errors due to sample variations

In XRF, the irradiated volume of the sample that is analyzed depends on the density of the sample and the energy (penetrating power) of the analyte element X-rays. For most solid samples this can be ignored. However, for low-density liquid samples it becomes an important factor contributing to the accuracy of analysis, particularly for heavier elements.

Oil-Trace corrects for the varying volume of the fluorescence volume geometry (FVG). 

Compliancy with international norms

Oil-Trace supports compliancy with international ASTM and ISO norms and methods. It is particularly useful for the analysis of fuels, biofuels and mixtures, e.g. B20, E5, E85, additives in lubricating oils and wear metals in used oils, S, Ni, V and other restricted elements in crude oils and further refined oils, chlorine in compressor oils, metals in homogeneous liquid catalysts and troubleshooting for catalyst poisoning.

Unique setup samples

Oil-Trace is a package consisting of a suite of standards and specialized software. These standards are used to set up a base line calibration based on PANalyticals 5th generation FP model and reduce the need for many different reference materials. 

Specifications

Oil-Trace makes corrections to deal with varying finite thickness (FT) of samples as well as corrections to deal with varying fluorescence volume geometry (FVG) of samples. It is uses a 5th generation Fundamental Parameter (FP) model and deals with unmeasured light elements (C, H, O and N).