The Mysterious Case of the Musca Domestica Contaminant
Just when I thought I’d seen everything that could contaminate the background, a user sent me this picture! The Musca Domestica (common housefly) was attracted by certain aromatics used in testing and seems to have entered the measurement cell through an empty Hydro EV over a weekend. Unfortunately, it became wedged within the measurement cell and did not survive the experience.
The most important message from the Mastersizer training courses I give is to make sure you have a good background before you start. Laser diffraction is a background subtracted technique: the light scattered by the dispersant, the cell windows and the optical path is measured and subtracted from the signal read on the detectors. The background is measured each time a SOP or manual measurement is run.
The background should be stable and
- Detector 1 should be less than 100
- Detector 20 should be less than 20
- The profile should be a decreasing curve as shown below.
The measurement can only be as good as the sample preparation and the background. If the background is poor, results are likely to be poor also. The most common causes of poor backgrounds are not allowing the background to stabilize, contamination, and material on the cell windows. I now need to add houseflies to my contamination examples.
See a complete measurement in the Mastersizer 3000 with Hydro EV accessory in this video
Related content:
- To learn about method development in liquid dispersions see this webinar “Developing wet or liquid dispersion methods using laser diffraction“
- Read this application note focusing on “Wet or liquid dispersion method development for laser diffraction particle size distribution measurements“
Cet article a peut-être été traduit automatiquement
{{ product.product_name }}
{{ product.product_strapline }}
{{ product.product_lede }}

