Accurate Analysis of NMC Cathodes and Precursors Using Revontium XRF Spectrometer

Accurate XRF Analysis of NMC Battery Cathodes and Precursors

Lithium nickel manganese cobalt oxide, commonly known as Li-NMC or NMC, is an important cathode material used in lithium-ion batteries for electric vehicles, portable electronics, and energy-storage applications. The concentrations of nickel, manganese, and cobalt influence critical aspects of cathode performance, including energy density, stability, safety, and operating lifetime. Reliable elemental analysis is therefore essential throughout battery-material development and production.

This application note examines how energy-dispersive X-ray fluorescence, or EDXRF, can support quantitative analysis of NMC cathodes and precursor materials. XRF offers a rapid and straightforward alternative to techniques such as inductively coupled plasma analysis. Its combination of precision, accuracy, and limited analytical complexity makes it particularly relevant to routine quality assurance, raw-material qualification, research and development, and process control.

Why Calibration Materials Matter for NMC Analysis

Standardless XRF can provide rapid elemental screening and semi-quantitative results. Accurate production and quality-control measurements, however, require calibration using appropriate reference materials. Battery materials present a practical calibration challenge because suitable standards have historically been limited.

The method described in the application note uses a set of 12 synthetic NMC certified reference materials developed specifically for XRF calibration with fused beads. These CRMs cover nickel, manganese, cobalt, and selected minor elements across concentration ranges relevant to NMC materials. They may also support the analysis of NCA, LCO, LMO, and associated precursor materials.

The reference materials are prepared gravimetrically from high-purity chemicals and provide metrological traceability and compliance with ISO 17034. A fusion recipe and XRF method template are also included. These resources help laboratories establish a controlled and repeatable calibration workflow rather than relying only on standardless estimates.

Fused-Bead Preparation and Revontium EDXRF Measurement

Calibration standards and a validation CRM were prepared as 32-mm fused beads using lithium-borate fusion. The preparation used a 1:10 sample-to-flux ratio and a fusion temperature of 1100 °C. Fusion is recommended when high accuracy and repeatability are required because it can provide homogeneous specimens for XRF measurement. The CRMs can also be used to prepare secondary pressed-pellet standards.

Measurements were performed using a Revontium EDXRF spectrometer equipped with an Ag-anode X-ray tube, primary and secondary filters, four simultaneous high-resolution silicon drift detectors, a spinner, and an automatic sample changer. The measurements were conducted under vacuum with an oil-free dry pump and did not require helium.

Each specimen was measured for 4.5 minutes, corresponding to a throughput of approximately 13 samples per hour. Laboratories can adjust measurement time according to their required precision and analytical scope. Calibration curves were established for NiO, CoO, MnO, and SO3, linking measured XRF intensities with certified concentrations.

What the Full Application Note Covers

  • The concentration ranges covered by the NMC certified reference materials
  • The Revontium EDXRF instrument configuration and measurement conditions
  • The lithium-borate fused-bead preparation procedure
  • Calibration results for major and minor components
  • Same-day repeatability results from ten replicate fused beads
  • Method stability and precision results collected across three days
  • Comparison of measured averages with certified concentrations and ISO Guide 35-based allowable differences

The complete method was validated using BAM-S014, a certified Li-NMC 111 cathode material. The evaluation considered both the bead-making process and XRF measurement, providing a practical view of whole-method performance rather than instrument precision alone.

Register for full access to review the complete calibration, precision, and validation data and learn how Revontium EDXRF with NMC-specific CRMs can support efficient routine elemental analysis of battery cathodes and precursors.

Introduction

Lithium nickel manganese cobalt oxide (Li-NMC) cathodes are widely used in lithium-ion batteries for electric vehicles, portable electronics, and energy storage. The concentrations of Ni, Mn, and Co strongly influence cathode performance, including energy density, stability, safety, and lifetime. Accurate elemental analysis is therefore essential for raw-material qualification, process control, R&D, and quality assurance.

XRF provides a rapid and simple alternative to techniques such as ICP, offering high precision and accuracy with minimal analytical complexity. This makes it well suited to routine analysis of cathode and precursor materials. This application note demonstrates quantitative NMC analysis using a Revontium EDXRF spectrometer, NMC-specific certified reference materials (CRMs), and lithium-borate fused-bead preparation.

NMC Certified Reference Materials

Standardless XRF provides rapid elemental screening and semi-quantitative results. For accurate production and quality-control measurements, calibration with appropriate reference materials is required. To address the limited availability of battery-material standards, Malvern Panalytical has developed 12 synthetic NMC CRMs specifically for XRF calibration using fused beads.

[AN260907-image1.png] AN260907-image1.png

Table 1: Minimum and maximum concentrations of NCM CRMs

Li (wt%)*Mn (wt%)Co (wt%)Ni (wt%)Al (wt%)Ca (wt%)Zr (wt%)Na (wt%)S wt(%)
Lowest cal. point **5.703.003.0010.0000000
Highest cal. Point9.0027.0027.0055.002.000.102.001.000.40

* Lithium or Lithium oxide cannot be measured directly by XRF instruments, but they are added to the composition to simulate battery cathode mixes.

** The lowest calibration point should not be considered as the minimum concentration that can be reported. Instead, the Limit of Quantification (LOQ) is used for this purpose. LOQ depends on sample preparation, XRF instrument, measurement conditions, and measurement time.

The CRMs are prepared gravimetrically from high-purity chemicals, providing metrological traceability and compliance with ISO 17034. The package also includes a fusion recipe and XRF method template.

The CRMs cover Ni, Mn, Co and selected minor elements across the concentration ranges relevant to NMC materials. They can also be used for NCA, LCO, LMO and their precursor materials. Fusion preparation is recommended for high accuracy and repeatability, while the CRMs can also be used to prepare secondary pressed-pellet standards.

Instrumentation and Sample Preparation

Measurements were performed on a Revontium EDXRF spectrometer equipped with a 50 W, 5 mA, 60 kV Ag-anode X-ray tube, primary and secondary filters, four simultaneous high-resolution silicon drift detectors, a spinner, and an automatic sample changer. Measurements were performed under vacuum using an oil-free dry pump, without the need for helium.

Calibration standards and the validation CRM were prepared as 32-mm fused beads via lithium-borate fusion. A sample-to-flux ratio of 1:10 was used, with fusion performed at 1100 °C using an Eagon 2 automatic fusion machine. The complete cold-to-cold cycle required approximately 30 minutes.

Measurement and Calibration

Each specimen was measured for 4.5 minutes, providing a throughput of approximately 13 samples per hour. Measurement time can be adjusted according to the required precision and analytical scope.

Calibration curves for NiO, CoO, MnO, and SO3 showed excellent correlation between measured XRF intensities and certified concentrations, demonstrating the suitability of the calibration materials and analytical method.

[AN260907-figure1.png] AN260907-figure1.png
Figure 1: Calibration plot of NiO in NCM CRM standards prepared as fused beads.
[AN260907-figure2.png] AN260907-figure2.png
Figure 2: Calibration plot of CoO in NCM CRM standards prepared as fused beads.
[AN260907-figure3.png] AN260907-figure3.png
Figure 3: Calibration plot of MnO in NCM CRM standards prepared as fused beads.
[AN260907-figure4.png] AN260907-figure4.png
Figure 4: Calibration plot of SO3 in NCM CRM standards prepared as fused beads.

Precision and Validation

The complete method was validated using BAM-S014, a certified Li-NMC 111 cathode material. Ten replicate fused beads measured on the same day showed close agreement with the certified concentrations, as shown in Table 2.

Table 2: Precision and validation results of the whole method including the bead-making process on the same day. The smaller the RMS value, the better the precision.
Sample IDNiO (wt%)CoO (wt%)MnO (wt%)SO3 (wt%)
BAM-S014_repeat_bead_0119.50519.62517.9740.1450
BAM-S014_repeat_bead_0219.63419.77718.0790.1498
BAM-S014_repeat_bead_0319.77819.86718.1750.1482
BAM-S014_repeat_bead_0419.73819.83618.1450.1490
BAM-S014_repeat_bead_0519.67219.77418.1040.1474
BAM-S014_repeat_bead_0619.69419.81018.1170.1490
BAM-S014_repeat_bead_0719.78219.90218.1850.1482
BAM-S014_repeat_bead_0819.69919.78818.1030.1502
BAM-S014_repeat_bead_0919.69019.82018.1220.1486
BAM-S014_repeat_bead_1019.65919.78718.1160.1558
Measured average conc.19.68519.79918.1120.149
RMS of average conc.0.0790.0740.0590.003
Certified conc.19.76019.80018.2200.142
RMS of certified conc.0.1300.1200.1400.007
RMS between labs0.2100.2000.2500.012

The method was also evaluated over three days. As shown in Table 3, the measured averages remain close to the certified values, with differences well within the allowable limits calculated according to ISO Guide 35.

Table 3: Precision and validation results of the whole method including bead making process, spread over 3 days. The smaller the RMS value, the better the precision.
Sample IDNiO (wt%)CoO (wt%)MnO (wt%)SO3 (wt%)
BAM-S014_day 1_bead 119.67119.77318.0870.1422
BAM-S014_day 1_bead 219.74919.87318.1570.1438
BAM-S014_day 2_bead 119.80119.92118.2080.1454
BAM-S014_day 2_bead 219.64819.78318.0930.1426
BAM-S014_day 2_bead 319.60819.73218.0560.1494
BAM-S014_day 3_bead 119.72419.85718.1340.1438
BAM-S014_day 3_bead 219.84619.96618.2340.1490
Measured average conc.19.72119.84318.1380.145
RMS of average conc.0.0850.0850.0660.003
Certified conc.19.76019.80018.2200.142
RMS of certified conc.0.1300.1200.1400.007
RMS between labs0.2100.2000.2500.012
Absolute difference0.0390.0430.0820.003
Allowed difference*0.1550.1470.1550.007

*Allowed differences between average concentration and certified concentration, calculated according to the ISO Guide 35 requirements.

This demonstrates good repeatability and stability of the complete sample-preparation and measurement procedure.

Conclusion

Revontium EDXRF provides accurate, precise, and repeatable quantitative analysis of NMC cathode and precursor materials prepared as fused beads. The combination of NMC-specific CRMs, reproducible fusion preparation, and rapid XRF measurement provides a robust solution for battery-materials R&D, process control, and quality assurance. With a 4.5-minute measurement time and throughput of approximately 13 samples per hour, Revontium enables efficient routine elemental analysis of NMC materials.