Solvent Peaks – How to get rid of the buffer/salt/additive DLS contribution?
Have you ever seen a peak at small size that you think just should not be there?
This ‘solvent peak’ can happen under certain conditions, and it is usually due to the signal from the buffer component in the sample. Ultimately, the high sensitivity of today’s dynamic light scattering (DLS) systems like the Zetasizer Nano has increased the observable range both in terms of concentration and size, thus ever smaller molecules can be detected (with the caveat that the required concentration has to be increased in order to see them, as the laws of physics still apply!).
A recent solvent peak query from one of our customers reads as follows:
“I just read your comment on Arg for stabilization on Malvern’s website. I hope you can help me with a concern we have with Arg. We are currently working with virus-like particles (VLP) in PBS buffer with 0.2 M L-Arg added. We consistently find a peak at about 0.8 nm that accounts for 99% of the species either in %vol or %number. The peak is present in the buffer with or without protein. The VLP in PBS without L-Arg does not show the peak. Our equipment is a Zetasizer microV, and the samples are measured at 20ºC. Have you ever seen this peak? Any hints on how to get rid of it?”
We have seen the scattering from Arginine before as well, which can also manifest itself as a solvent peak. The reason this ghost peak appears is that the Arg molecules do scatter enough light to be seen by the detector, and the fluctuations are slow enough to appear as a correlation function.
The peak is real in the sense that there actually is something of that size in your sample (for example, see 1nm sucrose peak). Buffer components that may cause these contributions are additives such as salts, ammonium sulfate, glycerol, mannose, trehalose, glucose, amino acids. We also refer to the small size contribution with these alternate names: solvent peak, ghost peak, salt peak, amino acid peak, sugar peak, in addition to “buffer peak”.
How to eliminate the buffer peak
The buffer peak can be eliminated by changing the display range of the analysis. In order to access this parameter, just edit the record by
- Highlighting the record

- Right-click Edit Result…
- Data Processing
- Configure
- Display Range : Lower limit
….and change the lower limit to 1 nm, for example. See the effect of this change in the graph below, displaying the DLS result from a monoclonal antibody prepared in a buffer with Arginine (red before editing, green after editing). The volume distribution then shows that the relevant data is still there.
Alternatively, if you have the flexibility to try this, increasing the concentration of your sample particles or molecules of interest may suppress the buffer peak. Alternatively, you could reduce the concentration of the ‘offending’ buffer component – but that is not usually very practical!
Two extra tips to eliminate the buffer peak (aka solvent peak or ghost peak)
If you encounter this more than occasionally, here are two extras.
- Bonus Tip#1: The above trick also works for editing multiple records at the same time: just highlight all desired records, right-click edit, and place a checkmark in the box next to “Apply all analysis settings to selected measurements” under Data Processing – Measurement collection. Then re-analyze all selected records with the revised display range.
- Bonus Tip#2: The display range can of course be part of the SOP setup. So once you identify this concern, you can modify the SOP to adjust the display range for further samples. The resulting record then automatically takes these settings into account, for all samples analyzed using this SOP (as an aside: no need to worry, you can always edit data and reanalyze without any limits, if that’s a concern).
Finally, a higher-resolution distribution algorithm like the multiple narrow mode or the protein analysis mode may increase the sensitivity towards detecting buffer peaks. The general purpose (normal resolution) mode would be the preferred and easiest choice to circumvent the issue. The different models are discussed in the technical note “Zetasizer Nano Analysis Methods Explained“.
Previously
- What is a narrow band filter?
- 50,000 Zetasizer references
- What protein sample concentration is good for DLS?
If you have any questions, please email me at ulf.nobbmann@malvern.com. Thanks! While opinions are generally those of the author, some parts of this blog on solvent peaks may have been modified by our editorial team.
這篇文章可能已自動翻譯
{{ product.product_name }}
{{ product.product_strapline }}
{{ product.product_lede }}
