Zetasizer WT
On-line zeta potential analyzer for water treatment processes
The Zetasizer WT is a fully automated system that delivers lab quality on-line zeta potential measurements. This enables real-time monitoring of flocculation process performance and efficient charge neutralization. On-line charge measurements you can trust are finally here!
Overview
The Zetasizer WT combines Malvern Panalytical’s industry leading Electrophoretic Light Scattering technology with two decades of expertise with in-process measurements to produce a dedicated on-line zeta potential analyzer for water treatment plants.
No more guess work – accurate and reliable results in [mV] provide a clear range for dosing control.
Respond to changes in raw water before they cause filtration problems!
Reduce chemical use to optimum for forming a stable floc. Overdose no more!
- Completely automated zeta potential analysis with results in minutes.
- Results sent to SCADA via OPC, Modbus or 4-20mA
- Minimal maintenance
- Easy measurement of grab samples on the same system.
- HMI for local control and setup
- On board computer for autonomous operation
- Sample flow and temperature sensing
How it works
The Zetasizer WT connects to a sample stream coming from the plant after coagulants have been added and mixed into the water. The system uses a pump to flow a sample through the analyzer`s flow cell where zeta potential measurements are made. Sample is then pumped into waste or returned back into the process and replaced with a fresh sample in a continuous, automatic cycle.
Automation and analysis is controlled by an on-board industrial computer running Malvern Panalytical Link II software.
Measurement results are displayed on a local HMI screen as well as exported into plant SCADA via OPC, Modbus TCP/IP or wired 4-20mA connections.
Results are used by treatment plant operators or a plant control system to adjust coagulant dose required to keep zeta potential in a pre-determined range – typically between zero and a slight negative charge.
Zetasizer WT also allows easy measurement of grab samples from other areas in the process, effectively doubling as a laboratory system.
Specification
Sample requirement
- Sample flow rate:
- 100 - 350 LPH (fast loop)
- Sample flow rate:
- 0.5 - 2 LPM (measurement)
System
- Measurement type:
- Zeta potential
- Measurement principle:
- Laser Doppler micro-electrophoresis
Measurement
- Result output frequency:
- 3 minutes typical, programmable
- Operation:
- 24hr continuous
- Range of zeta potential measurement:
- +/- 500 mV (+/- 90 mV on HMI interface)
General
- Size range:
- 3.8 nm to 100µm
- Enclosure rating:
- IP54, indoor use. Limited capability to NEMA 4X
- Power:
- 90 - 132 VAC, 180 - 264 VAC, 50 - 60 Hz, 6 A
- Dimensions (W, D, H):
- 855 x 686 x 1720mm
- Weight:
- 160 Kg
- Compressed air purge:
- 1.5LPM, 6-10 bar, ISO 8573-1:2010 [0.7.2] (optional)
Operating environment
- Temperature range:
- 0°C to 40°C
- Humidity:
- 5-95% non-condensing
- Maximum sample pressure:
- 6 bar
Accessories
Zetasizer WT
Disposable folded capillary cell (DTS1070)
Eliminate sample cross contamination between zeta potential measurements using a completely disposable cell.
The zeta potential of colloids and molecules is extremely sensitive to contamination from components of a previous sample used in a measurement cell. The use of a disposable cell completely eliminates this potential contamination issue.
The folded capillary cell provides the benefits of the accuracy of measurement of a capillary cell with the ease of use of a disposable cell.
Although, disposable, including the electrodes, for low conductivity samples each cell can be used for hundreds of measurements if required.
The cell can be used with the autotitrator, where multiple measurements are routinely done over a wide pH range.
Support
Protect your investment and ensure optimized performance at all times with Malvern Panalytical's service plans.
When you purchase a Malvern Panalytical product we understand that this is just the first stage of a working relationship that will last for the lifetime of the instrument. Depending on your needs, Malvern Panalytical will provide the support for your business.
Help Desk
- Telephone and email support.
- Operators who take responsibility for your call and ensure your query is answered.
Training
Software downloads
Malvern Panalytical’s service options- choose the right plan for you:
Malvern Panalytical Platinum Plan
For laboratories where maximizing instrument up-time is critical to its daily performance. Our highest level of response and the 'all inclusive' price plan takes care of those unexpected repair bills should the unfortunate occur.
Malvern Panalytical Gold Plan
Enhance your laboratory productivity by maintaining instrument efficiency. Still receive that priority response, combined with specialist technical and software support, we'll keep your instrument performance on track to deliver.
Malvern Panalytical Bronze Visit
For laboratories looking to optimize instrument performance, regular preventative maintenance is essential.
Platinum Plan | Gold Plan | Bronze Visit | |
---|---|---|---|
Annual PM/PV Test | yes | yes | no |
Telephone / Email Support | yes | yes | no |
Priority Response | yes | yes | no |
Emergency Breakdown visits* | yes | yes | no |
Parts Included | yes | no | no |
Technical & Software Support | yes | yes | yes |
IQ/OQ (Pharma)*** | yes | yes | yes |
*including labour & travel costs***available at an additional cost