Size quality report for the Zetasizer Nano

Twelve indicators of result quality are  combined to give a quality factor which is displayed on every report  to simplify data interpretation.  A separate size quality report  provides more detail, and this technical note expands on this.

Introduction

The quality of data obtained from a dynamic light scattering (DLS) measurement is paramount to the reliability of the result obtained. To simplify the interpretation of the data from a dynamic light scattering measurement, a size quality report has been developed for the Zetasizer Nano range of instruments.

This size quality report is displayed by default as one of the tabs in the size workspace. Several tests are performed on a selected record and if any of the tests fall outside specified limits, a warning message is displayed together with advice of possible reasons for the warning, and a result quality ‘Refer to quality report’ is displayed in green in the sample report. If none of the tests fail, a “Result Meets Quality Criteria” message is displayed in the Size Quality Report, and a result quality ‘Good’ is displayed in green in the sample report.

The size quality report incorporates 12 tests on any selected record. Table 1 is a summary of the tests incorporated and lists descriptions and details of the tests, the warning message that is displayed, possible reasons for the warning message and actions that might remedy it.

This technical note uses example results to illustrate the warning messages and provides details of possible actions of how to improve the results.

Table 1: Summary of the tests incorporated into the size quality report listing descriptions and details of the tests, the warning message that is displayed, possible reasons for the warning message and possible actions that might remedy it
Test NumberTest DescriptionTest DetailsWarning MessagePossible Reasons
for Warning Message
Possible Actions
1Check appropriate lower size analysis limit used in the distribution analysisIs the z-average diameter < lower analysis limit?z-average is smaller than lower size analysis limitThe wrong lower size analysis limit used in the distribution analysis settings. Only applicable to research softwareEdit the lower size analysis settings in the research software
2Check appropriate upper size analysis limit used in the distribution analysisIs the z-average diameter > upper analysis limit?z-average is larger than upper size analysis limitThe wrong upper size analysis limit used in the distribution analysis settings. Only applicable to research softwareEdit the upper size analysis settings in the research software
3Check appropriate lower size display limit usedIs the z-average diameter < lower display limit?z-average is smaller than lower display limitThe wrong lower display limit used.Edit the lower display settings
4Check appropriate upper size display limit usedIs the z-average diameter > upper display limit?z-average is larger than upper display limitThe wrong upper display limit used.Edit the upper display settings
5Check polydispersity index valueIs the polydispersity index value > 1?Polydispersity index is very highSample is very polydisperse and
may not be suitable for DLS measurements
Sample may not be suitable for DLS technique
Sample contains large particles/ aggregates/ dustRemove large particles/ aggregates/ dust by filtration or centrifugation
Wrong measurement position used (applicable only to Nano S/ZS/ZSP)Seek for optimum measurement position should be selected
6Check correlation function intercept valueIs the correlation function intercept <0.1 or >1.0?Correlation function intercept out of rangeSample concentration too high (multiple scattering)Dilute sample and re-measure
Sample concentration too lowIncrease sample concentration and re-measure
Sample fluorescenceUse narrow band filter option
User different laser wavelength
Sample absorbance (coloured sample)User different laser wavelength
Wrong measurement position selected (Nano S/ZS/ZSP only)Allow instrument to seek for optimum measurement position
Sample contains very large particles (baseline definition problems) and may not be suitable for DLS measurementsRemove large particles/ aggregates/ dust by filtration or centrifugation
7Check first point selection for multimodal analysisIs the first correlation point >3 and the z-average diameter <5nm?Check first correlation point selection for multimodal analysisOnly applicable to research software and is a check that the first correlation point selection for the distribution analysis is suitable for the particle size obtainedEdit the first point selection for the distribution analysis and re-analyze the record
8Check the in range valueIs the in range value less than 90%The in range figure is lowPresence of large or sedimenting particlesRemove large particles/ aggregates/ dust by filtration or centrifugation
Sample fluorescenceUse narrow band filter option
User different laser wavelength
Sample absorbance (colored sample)User different laser wavelength
9aCheck for low mean count rateIs the mean count rate <20 kcps?Count rate is out of rangeAttenuator not set to automaticAllow instrument to seek for optimum attenuator position
Sample concentration too lowIncrease sample concentration and re-measure
Sample absorbance (colored samples)User different laser wavelength
Sample is not stable during measurementSample may not be suitable for DLS technique
9bCheck for high mean count rateIs the mean count rate >1000 kcps?Count rate is out of rangeAttenuator not set to automaticAllow instrument to seek for optimum attenuator position
Sample contains large particles/aggregates/dustRemove large particles/ aggregates/ dust by filtration or centrifugation
Sample is not stable during measurementSample may not be suitable for DLS technique
10Check sufficient data collectedIs the total number of photons collected <10000K?Insufficient signal collectedMeasurement duration not set to automaticSet instrument to automatic duration
Attenuator not set to automaticAllow instrument to seek for optimum attenuator position
Filter factor not set to default (50%).
Only applicable to research software
Edit filter factor setting in the research software
11Check the cumulants fit errorIs the cumulants fit error >0.005?Cumulant fit error highData quality too poor for cumulant analysisSample may not be suitable for DLS technique
Sample too polydisperse for cumulant analysisSample more suitable for distribution analysis
Inappropriate cumulant analysis settings in research softwareEdit cumulant settings in research software
12Check the distribution fit errorIs the distribution fit error >0.005?Multimodal fit error highData quality too poor for distribution analysisSample may not be suitable for DLS technique
Sample too polydisperse for distribution analysisSample may not be suitable for DLS technique
Inappropriate distribution analysis settings in research softwareEdit distribution settings in research software

The (R) after a test indicates that the test is only applicable when using the optional research software

Test 1 (R)

Warning message

Z-Average is smaller than lower size limit

Is the z-average diameter < lower analysis limit?

This test is only applicable to the research version of the Nano software where the distribution analysis limits have been manually selected incorrectly. Figure 1 shows the intensity size distribution obtained for a protein sample with a z-average diameter of 8.9nm. The lower distribution analysis limit has been manually set to 10nm in the research software. This has given “z-average is smaller than lower size analysis limit” warning message in the size quality report. To rectify the problem, the record needs to be selected in the Record View and edited (by clicking on Edit, Edit Result). This action brings up the Edit Result window. The Update Size Model option needs to be selected on the Size Analysis tab. After the appropriate distribution analysis model has been selected (General Purpose, Multiple Narrow Modes or CONTIN), the Size Classes tab should be accessed (figure 2). The Lower Limit needs to be changed either to the appropriate value or reset to the Default value of 0.4nm.

Figure 1: An intensity size distribution of a protein sample with a z-average diameter of 8.9nm where the lower distribution analysis limit has been manually set to 10nm in the research software
mrk693 fig1
Figure 2: An appropriate value for the Lower Limit in the Size Classes tab needs to be selected.
mrk693 fig2

A new record will be calculated with these new settings implemented and will be displayed in the Record View. Figure 3 shows the intensity size distribution obtained for the edited record displayed in figure 2. The Size Quality Report now displays a “Result Meets Quality Criteria” message.

Figure 3: The intensity size distribution obtained for the edited record displayed in figure 2.
mrk693 fig3

Test 2 (R)

Warning message

Z-Average is larger than upper size analysis limit

Is the z-average diameter > upper analysis limit?

This test is only applicable to the research version of the Nano software where the distribution analysis limits have been manually selected incorrectly. This test is similar to that discussed in test 1 and any warning message can be rectified in the same way as discussed above (figure 2). The only difference being in the value used as the upper analysis limit.

Test 3

Warning message

Z-Average is smaller than lower display limit

Is the z-average diameter < lower display limit?

This test ensures that a suitable lower display limit is selected for the result. A warning message for this test would only be obtained if the user had manually set the lower display limit to an inappropriate value. Figure 4 shows the intensity size distribution of the protein sample used above as an example where the lower display limit has been set to 15nm. The left hand side of the distribution has been cut off at the lower display limit that has been set. This has given the warning message “z-average diameter is smaller than lower size display limit” in the size quality report. To correct this problem, the lower display limit needs to be edited in the record to an appropriate value.

Figure 4: The intensity size distribution obtained for a protein sample with a z-average diameter of 8.9nm and a lower display limit set to 15nm.
mrk693 fig4

After selecting the record and clicking on Edit, Edit Result, the Size Analysis Settings tab needs to be selected in the Edit Result window. Selecting the Update Analysis Model allows the Advanced button to be accessed (figure 5). Note that the Size Analysis Settings are much simpler in the normal software compared to the Research software (figure 2). Also note that the Advanced button displays a red exclamation mark indicating that one of the settings is not at default value.

Figure 5: Selecting the Update Analysis Model on the Size Analysis Settings tab allows the Advanced button to be accessed
mrk693 fig5

Clicking on the Advanced button gives access to the Advanced Analysis Settings (figure 6). This allows the Display Size Limits to be modified to appropriate values. Note that the default values for the lower and upper display limits are 0.6 and 6000nm respectively.

##fig6

Figure 6: The Advanced Analysis Settings tab allows the Display Size Limits to be edited
mrk693 fig6

A new record will be created with the edited settings in the Record View. The intensity size distribution obtained for the protein sample using default display limit settings is shown in figure 3. The Size Quality Report gives a “Result Meets Quality Criteria” message.

Test 4

Warning message

Z-Average is larger than upper display limit

Is the z-average diameter > upper display limit?

This test is similar to that discussed in test 3 and any warning message can be rectified in the same way as discussed above (figures 5 and 6).

Test 5

Warning message

Polydispersity index is very high

Is the Polydispersity Index value > 1?

The polydispersity index (PDI), together with the z-average diameter, is calculated from the cumulants analysis as defined in ISO22412:2008 and is a dimensionless estimate of the width of the distribution, scaled from 0 to 1.

A PDI value of 1 indicates that the sample has a very broad size distribution and may contain large particles or aggregates that could be slowly sedimenting. If this is the case, the sample may not be suitable for a DLS measurement. Figure 7 shows a correlation function obtained from a sample consisting of a broad size distribution containing large, sedimenting particles (note the noisy baseline). This sample gave a PDI value of 1 and is a good example of a sample not suitable for the DLS technique. Note the intercept value is greater than 1.0 and further discussion of this will be given in test 6 below. Often, when a sample gives a PDI value of 1.0, cumulants and distribution fit errors are also obtained and these will be further discussed in tests 11 and 12 below.

Figure 7: A correlation function from a sample consisting of a broad size distribution containing large, sedimenting particles
mrk693 fig7

Another possible reason for obtaining a warning message with test 5 is that the wrong measurement position has been selected in a Zetasizer Nano S or ZS instrument. The default option is to allow the software to automatically seek for optimum position. However, the measurement position can be manually selected. The default option of seeking for the optimum measurement position is accessed by clicking the Advanced Button on the Measurement Settings window either when creating an SOP or setting up a Manual Measurement (figure 8).

Figure 8: Clicking the Advanced button on the Measurement Settings window allows access to the default option of seeking for the optimum measurement position
mrk693 fig8

Test 6

Warning message

Correlation function intercept out of range

Is the correlation function intercept <0.1 or > 1.0?

The intercept value is the signal/ background ratio and the value will vary according to the sample and instrument optical configuration. Appropriate values are between 10% and 100% (i.e. 0.1 to 1.0). A lower than expected value may indicate too high or too low a sample concentration, absorption or fluorescence of the sample. Too high a sample concentration may lead to multiple scattering effects that will reduce the intercept value. The NIBS configuration of the Zetasizer Nano S/ZS allows for the path length of the laser beam to be minimized, which, in turn, will minimize any multiple scattering present. The wrong measurement position may have been manually selected resulting in multiple scattering effects. The default setting of allowing the software to automatically seek the optimum measurement position can be accessed as discussed in Test 5 above (figure 8).

Figure 9 shows a correlation function obtained from a sample exhibiting fluorescence and shows an intercept value of less than 0.1. Improved data may be obtained by using the narrow band filter option available for the Zetasizer Nano series. If this does not improve the data quality sufficiently then an instrument with a different wavelength of laser would be required.

Figure 9: A correlation function obtained from a sample exhibiting fluorescence. Note the very low intercept value of 0.011 obtained
mrk693 fig9

The correlation function shown in figure 7 above was obtained from a sample containing large, sedimenting particles which cause interference in the measurement (number fluctuations). The presence of such large particles results in an ill-defined baseline and this causes the intercept value to be greater than 1.0. In addition, in range, cumulants and distribution fit error messages may also be obtained (see tests 8, 11 and 12 for further information). This sample will only be suitable for measurement using DLS if the large particles present can be suitably removed by, for example, filtration or centrifugation.

Test 7 (R)

Warning message

Check first correlation point selection for multimodal analysis

Is the first correlation point >3 and the z-average diameter <5nm?

This test is only applicable to the research version of the Nano software and checks whether the first correlation point used in the distribution analysis is suitable for the particle size obtained. A warning message for this test would only be obtained if the user had manually set the first correlation point to an inappropriate value with the z-average diameter less than 5nm. To correct this problem, the record needs to be selected in the Record View and edited (by clicking on Edit, Edit Result). This action brings up the Edit Result window. The Update Size Model option needs to be selected on the Size Analysis tab. After the appropriate distribution analysis model has been selected (General Purpose, Multiple Narrow Modes or CONTIN), the Points Selection section can be accessed on the chosen distribution analysis tab (figure 10). The First Channel of the Point Selection needs to be changed either to an appropriate value or reset to the Default value of 3. A new record will be calculated with these new settings implemented and will be displayed in the Record View.

Figure 10: The points selection section can be accessed on the chosen distribution analysis tab in the Edit Result window
mrk0693_figure10

Test 8

Warning message

The in-range figure is low

Is the in range value less than 90%?

The in range value is an indicator of the overall quality for good data obtained from a measurement. The maximum value achievable is 100% and most of the effects that conspire to ruin a measurement will reduce this figure.

The main reason for obtaining a warning message with this test is that the sample contains large, sedimenting particles. A typical correlation function from such a sample is shown in figure 7 above in which the sample is very polydisperse and the large particles present are causing baseline definition problems. This sample is not suitable for DLS measurement unless the large particles are removed by filtration or centrifugation.

Other possible reasons for low in range values are sample fluorescence or absorbance in which definition of the baseline may be problematic and/or the intercept value of the correlation function may be very low (see test 8 and figure 9 above).

Test 9a

Warning message

Count rate is out of range

Is the mean count rate <20kcps?

This test checks that sufficient scattering is being detected during a measurement.

There are various reasons that can account for this warning message, most of which are sample related. These include a sample concentration that is too low, a sample that shows strong absorbance of the laser beam, or a sample that is not stable during the measurement and may not be suitable for the DLS technique.

The Zetasizer Nano has an automatic attenuator that varies the intensity of the laser entering the sample cuvette. This subsequently varies the intensity of the scattered light detected. For samples that do not scatter much light, such as very small particles or samples of low concentration, the amount of scattered light being detected must be increased. The attenuator will therefore automatically allow more light through to the sample. For samples that significantly scatter light, such as large particles or samples at high concentration, the attenuator automatically reduces the amount of light that passes through to the sample. The attenuator in the Zetasizer Nano has 11 positions covering an attenuator range of 99.997% (position 1) to 0% (position 11). Therefore, if the attenuator position automatically selected is 11 (0% attenuation or 100% transmission) and the mean count rate is less than 20 kilo counts per second (kcps), it is indicative of very low sample concentration or strong sample absorbance. If the problem is low concentration, then the sample concentration needs to be increased if possible and re-measured. If the problem is strong absorbance, the only successful way to improve the results may be to use a different wavelength of laser.

Particles in a sample may be slowly dissolving, sedimenting or creaming during the measurement. In each of these cases, the mean count rate will reduce with time. Such an unstable sample may not be suitable for measurement by DLS. A more suitable solvent may prevent dissolution and hence stabilize the sample.

Removal of the sedimenting particles by filtration or centrifugation will improve the suitability of the sample to DLS measurement. Particle sedimentation will also give an in range warning message as discussed in Test 8 above.

The final reason for this warning message may be that the attenuator setting is not set to automatic and an inappropriate setting has been used where the attenuation of the laser beam is too high. It is recommended that the instrument be allowed to select the optimum attenuator position. This default option of selecting the optimum attenuator position is accessed by clicking the Advanced Button on the Measurement Settings tab either when creating an SOP or setting up a Manual Measurement (figure 11). Checking the Seek For Optimum Attenuator option in the Advanced – Measurement Positioning and Attenuation window ensures that a count rate as close to 300kcps will be used (if this is achievable).

Figure 11: Clicking the Advanced button on the Measurement Settings tab allows access to the default option of seeking for the optimum attenuator position
mrk693 fig11

Test 9b

Warning message

Count rate is out of range

Is the mean count rate >1000kcps?

This test is similar to 9a with the exception that it tests whether the scattering detected is too high.

One reason for a warning message is that an inappropriate attenuator setting has been used (too low an attenuation). This can be rectified in the same way as discussed in Test 9a above and shown in figure 11.

Other reasons for a warning message with this test are sample related. The sample may contain large particles that cause high levels of scattering and these will need to be removed by allowing them to naturally sediment with time or by filtering or centrifuging the sample. Alternatively the sample may be slowly aggregating with time causing an increase in the scattering intensity. If this is the case, the dispersion needs to be stabilized in some way (e.g. addition of surfactant, adjustment of pH etc).

Test 10

Warning message

Insufficient signal collected

Is the total number of photons collected <10000K?

This test checks that sufficient data has been collected. There are three possible reasons for obtaining a warning with this test and they are all related to incorrect settings in the software.

Firstly, the measurement duration may have been manually set to an inappropriate value. The default option of automatically allowing the software to determine the optimum measurement duration is recommended. This is done by selecting the Automatic Measurement Duration in the Measurement Settings window when setting up a Manual Measurement or creating an SOP (figure 12).

Figure 12: Automatic Measurement Duration can be selected in the Measurement Settings window when setting up a Manual Measurement or creating an SOP
mrk693 fig12

Secondly, the attenuator position may be set at an inappropriate value. This can be changed as discussed in Test 9a above (figure 11).

Finally, the Filter Factor may not be set to the default value of 50%. This is only applicable to the research software. Each size measurement taken by the Nano is divided into a series of sub runs. The default Filter Factor value of 50% means that only half of the sub runs taken are used in the final data analysis. The criterion for deciding which sub runs are selected is the mean count rate. The 50% of sub runs with the lowest mean count rates are the ones that are used in the analysis. This strategy is designed to minimize the effects of any dust contamination that may be present. If the Filter Factor has been manually set to an inappropriate value, insufficient data may have been used in the analysis.

A record can be edited so that more of the sub run data is used. This is done by selecting the relevant record in the Record View and edited (by clicking on Edit, Edit Result). This action brings up the Edit Result window. The Update Size Model option needs to be selected on the Size Analysis tab. Clicking on the Advanced button allows the Advanced Analysis Settings to be accessed (figure 13). Clicking on the Update Filtering option allows the Filter Factor to be changed to an appropriate value (figure 14). A new record will be calculated with these new settings implemented and will be displayed in the Record View.

Figure 13: The Advanced button can be clicked on the Size Analysis tab in the Edit Result window to give access to the Advanced Analysis Settings window (figure 14)
mrk693 fig13
Figure 14: Clicking on the Update Filtering option allows the Filter Factor to be changed to an appropriate value.
mrk693 fig14

Test 11

Warning message

Cumulants fit error high

Is the cumulants fit error > 0.005?

The cumulants analysis derives the z-average diameter and polydispersity index as defined in the International Standard ISO22412. The cumulants fit error is a measure of the quality of the cumulants fit to the measured data indicating whether the z-average diameter and polydispersity obtained for a particular measurement are reliable values.

A warning message is obtained with this test if the fit error is greater than 0.005. This may be due to the data quality being too poor or that the sample is too polydisperse for the cumulants analysis. Figure 7 shows a correlation function obtained from a sample consisting of a broad size distribution containing large, sedimenting particles. This is a good example of a sample not suitable for the DLS technique. A cumulants fit error message is obtained with this sample.

Figure 15 shows the correlation function and the intensity size distribution obtained for a bi-modal sample. The size quality report has given a cumulants fit error indicating that the z-average diameter (301nm) is not reliable in this example. The cumulants and distribution fits are shown in figure 16.

Figure 15: The correlation function and intensity size distribution (inset) obtained for a bi-modal sample
mrk693 fig15
Figure 16: The cumulants fit and distribution fit (inset) obtained for the bi-modal sample shown in figure 15.
mrk693 fig16

Whereas the cumulants fit is poor due to the bimodality of the sample, the distribution fit is excellent. Therefore, even though the z-average diameter for this record is not reliable, the intensity size distribution obtained from the distribution analysis is accurate.

Another reason for obtaining a cumulants fit error is inappropriate settings being used in the cumulants analysis. This is only applicable to the research software. If this is the case, the record can be edited by selecting the relevant record in the Record View and by clicking on Edit, Edit Result. The Update Size Model option needs to be selected on the Size Analysis tab in the Edit Result window. This allows the Advanced Settings of the Cumulants Analysis to be accessed (figure 17). After the settings are edited to more appropriate values, a new record will be created with the edited settings in the Record View.

Figure 17: The Advanced Cumulants Settings can be edited by clicking on Update Size Model on the Size Analysis tab in the Edit Result window
mrk693 fig17

Test 12

Warning message

Multimodal fit error high

Is the distribution fit error > 0.005?

The distribution fit error is a measure of the quality of the chosen distribution analysis fit to the measured data. In the standard Nano software, the distribution analyses that can be used are General Purpose and Multiple Narrow Modes. In the research version of the software, CONTIN and Discrete Exponential Analysis are also available.

A warning message is obtained with this test if the fit error is greater than 0.005. This may be due to the data quality being too poor or that the sample is too polydisperse suggesting that the sample may not be suitable for the DLS technique. An example of such a sample is given in figure 7. This sample will give warning

messages for the cumulants fit, the distribution fit, the in range test, the correlation function intercept test and the polydispersity index test.

Inappropriate settings in the chosen distribution algorithm can also cause a distribution fit error message. However, this is only applicable to the research software. More appropriate settings can selected by editing the relevant record in the Record View (by clicking on Edit, Edit Result). The Update Size Model option needs to be selected on the Size Analysis tab in the Edit Result window. After the appropriate distribution analysis model has been selected (General Purpose, Multiple Narrow Modes or CONTIN), more suitable settings can be edited in the chosen distribution analysis tab (figure 10) and a new record will be created in the Record View.

ログイン

Not registered yet? Create an account