Detecting oversized particles in battery materials

Using Size Sure and Hydro Insight on the Mastersizer 3000+

Battery cathode and anode materials are typically manufactured with a narrow particle size distribution. However, during the materials processing, fines, agglomerates, or contaminants may be introduced as well. While fines might enhance packing density and often desirable, large agglomerates and contaminants pose serious risks - causing local hotspots that degrade performance, shorten battery life, and compromise safety. Detecting these oversized particles is of great interest, but at the same time challenging as well, due to their low abundance in the electrode materials.

Laser diffraction is the industry standard for measuring particle size in battery materials. However, traditional systems often miss the presence of few large particles as their scattering signal is masked by the bulk population.

To address this, we explored the enhanced capabilities of the Mastersizer 3000+, by combining adaptive laser diffraction (Size Sure) and dynamic imaging (Hydro Insight) to improve the detection sensitivity and to visualize the individual particles for better insight.

Adaptive laser diffraction (Size Sure) on the Mastersizer 3000+ separates the overall diffraction signal into steady and transient states. While the steady state reflects the bulk material, transient events—caused by anomalous large particles—are independently recorded and analysed. This is made possible through the Size Sure technology, which captures 10,000 diffraction frames per second, offering unmatched temporal resolution.

Dynamic imaging (Hydro Insight) complements Size Sure by visually confirming the presence of large particles. The Hydro Insight accessory on Mastersizer 3000+ enables the capture of high-resolution images of individual particles for independent verification and classification (e.g. agglomerates vs. true contaminants).

Log in or sign up for free to continue reading.

Battery cathode and anode materials are typically manufactured with a narrow particle size distribution. However, during the materials processing, fines, agglomerates, or contaminants may be introduced as well. While fines might enhance packing density and often desirable, large agglomerates and contaminants pose serious risks - causing local hotspots that degrade performance, shorten battery life, and compromise safety. Detecting these oversized particles is of great interest, but at the same time challenging as well, due to their low abundance in the electrode materials.

Laser diffraction is the industry standard for measuring particle size in battery materials. However, traditional systems often miss the presence of few large particles as their scattering signal is masked by the bulk population.

To address this, we explored the enhanced capabilities of the Mastersizer 3000+, by combining adaptive laser diffraction (Size Sure) and dynamic imaging (Hydro Insight) to improve the detection sensitivity and to visualize the individual particles for better insight.

Adaptive laser diffraction (Size Sure) on the Mastersizer 3000+ separates the overall diffraction signal into steady and transient states. While the steady state reflects the bulk material, transient events—caused by anomalous large particles—are independently recorded and analysed. This is made possible through the Size Sure technology, which captures 10,000 diffraction frames per second, offering unmatched temporal resolution.

Dynamic imaging (Hydro Insight) complements Size Sure by visually confirming the presence of large particles. The Hydro Insight accessory on Mastersizer 3000+ enables the capture of high-resolution images of individual particles for independent verification and classification (e.g. agglomerates vs. true contaminants).

Case Study: Contamination Detection in LNCA Cathode Material

In this case study, we tested pure Lithium Nickel Cobalt Aluminium Oxide (LNCA) cathode material and five samples spiked with increasing concentrations of 42.7 µm spherical glass particles, named LNCA +G1 to LNCA + G5. All specimens were analysed on Mastersizer 3000+ platform using:

  • Dynamic Imaging (Hydro Insight)
  • Classical Laser Diffraction, and 
  • Adaptive Diffraction (Size Sure)

Samples were dispersed in water (with a small amount of IGEPAL CO-630 surfactant) using the Hydro MV dispersion unit.

1. Instrument Setup

The experimental setup (Figure 1) included:

  • Mastersizer 3000+
  • Hydro MV (automated dispersion unit with stirring, ultrasonication, and degassing) – minimizes bubbles that may interfere with large particles diffraction signal.
  • Hydro Insight (Dynamic imaging with standard resolution lens for analysing 1–300 µm size range)
[600x400-Hydro-Insight-MS3000.jpg] 637607439677621197JD.jpg
Figure 1. Instrument used in this case study – Mastersizer 3000+ with Hydro MV dispersion unit and Hydro Insight dynamic imaging.

2. Results

2.1 Dynamic Imaging with Hydro Insight:

500,000 particles were imaged per sample in ~20 minutes each. In pure LNCA, the software identified three particle populations: fines, regular particles, and agglomerates (Figure 2), with a Dv50 of 6.7 µm (Figure 3). 

[AN250716-figure2.png] AN250716-figure2.png
Figure 2. A total 500,000 particle images of LNCA sample were captured using the Hydro Insight accessory with the Mastersizer 3000+. Image analysis software sorted the particles into three populations: typical agglomerates (left), regular particles (middle) and fines (right).
[AN250716-figure3.png] AN250716-figure3.png
Figure 3. Volume distribution of particle sizes in LNCA sample obtained from the image of 500,000 particles using Hydro Insight, showing a Dv50 of 6.7 um.

Spiked samples (LNCA+G1 to LNCA+G5) revealed the presence of between 2 to 100 distinct glass particles per 500,000 particles, equating to 4 to 200 Particles per Million (PPM) detection. Image of the first two samples is shown in Figure 4.

[AN250716-figure4.png] AN250716-figure4.png
Figure 4. Spherical Glass particles detected in samples LNCA+G1 (left) and LNCA+G2 (right), in a total of 500,000 particles, using Hydro Insight accessory on Mastersizer 3000+.

2.2 Classical Diffraction:

Five replicate measurements of pure LNCA showed consistent results (Dv10 = 3.61 µm, Dv50 = 6.69 µm and Dv90 = 11.6 µm) as shown in Figure 5 and summarised in Table 1. These measurements also reveal the presence of fines at mean diameter of ~1µm, in addition to the regular particles with mode around 7 µm. 

[AN250716-figure5.png] AN250716-figure5.png
Figure 5. LNCA cathode material measured 5 times with Mastersizer 3000+ using classical laser diffraction without employing adaptive diffraction (Size Sure).
Table 1. Summary of 5 repeated classical laser diffraction measurements on LNCA sample.
Classical DiffractionSampleMeasurement #Dx (10) (μm)Dx (50) (μm)Dx (90) (μm)

LNCAMeasure13.646.711.7

LNCAMeasure23.576.6611.6

LNCAMeasure33.576.6611.6

LNCAMeasure43.646.711.6

LNCAMeasure53.646.711.7
Mean

3.616.6911.6
1×Std Dev

0.03560.02110.00763
1RSD (%)

0.9850.3150.0655

In the spiked samples, glass particles were detectable only in LNCA+G4 and LNCA+G5, where concentrations exceeded ~100 PPM (Figure 6). Lower concentration samples showed no glass particle peak.

[AN250716-figure6.png] AN250716-figure6.png
Figure 6. Classical laser diffraction data from pure LNCA and specimens spiked with glass particles measured on Mastersizer 3000+. The glass particle peak is visible in specimens LNCA+G4 and LNCA+G5. This peak is not visible at lower concentrations.

2.3 Adaptive Diffraction (Size Sure):

Size Sure maintained identical steady-state results to classical measurements but uniquely revealed transient signals from large particles even in low concentration samples. 

Steady and transient state data of 5 repeat measurements on pure LNCA sample is shown in Figure 7. The steady state data shows similar particle size distribution as in the classical measurement, including a small peak of fines.  The transient state data do not show any additional features in this sample.  Results of these measurements are summarized in Table 2.

[AN250716-figure7.png] AN250716-figure7.png
Figure 7. LNCA sample measured with Mastersizer 3000+ employing Size Sure adaptive laser diffraction to separate Steady state (left) and Transient state (right) results.
Table 2. Summary of 5 repeat adaptive diffraction measurements employing Size Sure adaptive laser diffraction on LNCA sample.
Size Sure MeasurementsSampleMeasurement #Dx (10) (μm)Dx (50) (μm)Dx (90) (μm)

LNCAMeasure13.646.711.6

LNCAMeasure23.646.711.6

LNCAMeasure33.646.711.6

LNCAMeasure43.646.711.6

LNCAMeasure53.646.711.6
Mean3.646.711.6
1×Std Dev0.0003260.0007140.00223
1RSD (%)0.008960.01070.0192

In the spiked specimens, the steady state data showed glass particle peaks only in samples LNCA +G4 and LNCA +G5 (Figure 8), just like the classical laser diffraction data with sensitivity ~ 100 PPM. 

[AN250716-figure8.png] AN250716-figure8.png
Figure 8. Steady state data measured with Size Sure adaptive laser diffraction on the Mastersizer 3000+ for pure LNCA and spiked samples. In the steady state data, the glass particle peak was only visible in samples LNCA+G4 and LNCA+G5. This peak is not visible at lower concentrations.

However, the transient state data showed glass particle peaks also in samples LNCA +G2 and LNCA +G3 (Figure 9), which had much lower glass particle concentrations, down to ~10 PPM.

[AN250716-figure9.png] AN250716-figure9.png
Figure 9. Transient state data measured with Size Sure adaptive laser diffraction on the Mastersizer 3000+ for pure LNCA and spiked samples. In the transient state data, the glass particle peak is visible in specimens LNCA +G2, and LNCA+G3 as well, in addition to LNCA+G4, and LNCA+G5. Only pure LNCA and LNCA +G1 did not exhibit a peak corresponding to glass particles.

Table 3 summarizes the results of all measurements in relation to detection sensitivity of contaminant glass particles. While Classical diffraction can detect 100 PPM and beyond, Size Sure adaptive diffraction could detect down to 10 PMM. Hydro Insight can in principle detect down to 1 PPM, albeit with longer measurement time.

Table 3. Summary of detection sensitivity of dynamic imaging (Hydro Insight), classical laser diffraction, and dynamic laser diffraction (Size Sure) to low levels of large contaminant particles.
Sample IDGlass Particles detected with Hydro Insight (Particles per Million - PPM)Large particle Signal detected with classical diffraction (Y/N)Large particle Signal detected with Size Sure laser diffraction (Y/N)
LNCA0NN
LNCA+G14NN
LNCA+G210NY
LNCA+G324NY
LNCA+G4100YY
LNCA+G5200YY

3. Conclusion

The Mastersizer 3000+ with Size Sure and Hydro Insight features significantly enhance the detection of low-abundance, oversized particles in battery materials:

  • Classical Laser Diffraction: Detection limit ~100 PPM (~ 2-minute measurement)
  • Size Sure: Detection limit ~10 PPM (~2-minute measurement)
  • Hydro Insight Dynamic Imaging: Detection limit ~1 PPM, but longer measurement time (~40 minutes for 1 million particles imaging)

This powerful combination enables advanced QA/QC with large particle detection, improving battery safety, yield, lifetime, and performance.

ログイン

まだ登録されていませんか? アカウントを作成する