The Viscotek GPC Heated Column Ovens can hold up to five analytical GPC/SEC columns in an anodized aluminum compartment that is resistant to corrosion by most types of mobile phases.

A column oven can improve a GPC/SEC measurement by helping to stabilize detector baselines and improve separation efficiency.The systems includes a ‘ready’ light to indicate that it has reached the required temperature.

The higher temperature unit can also accommodate a sample loop for temperature equilibration before injection.

動作

In a typical GPC/SEC system, the column oven is placed between the pump and the detectors, so that the columns are maintained at a stable temperature as the rest of the system. This helps to stabilize the baseline, improve separation and give better signal to noise ration during the analysis.

仕様

一般

動作環境:
Ambient to 100°C or 150°C
温度精度:
±2°C
温度再現性:
±1°C
温度安定性:
±0.2°C
温度安全性:
118°C
推奨流量:
Up to 8mL/min
カラム数:
Up to five 8mm x 300 mm analytical columns
電源:
100-120v single fused, 220-240v double fused
最大消費電力:
120W
寸法(W x D x H):
12.7, 49.5, 22.3cm
重量:
9.5 kg