MPT-3 自動滴定装置

pH滴定を自動化

MPT-3はゼータサイザーアドバンスシリーズの効率性の高いアクセサリーで、ゼータ電位や粒子径のpH依存性確認の測定を自動化するように設計されており、材料がpHの関数としてどのように機能するかを理解するのに役立ちます。    

 pH変化による材料のゼータ電位や粒子径を理解することは非常に重要です。

  • 材料安定性評価
  • 凝集傾向評価
  • 凝集剤評価 
  • 等電点測定 
  • 保管条件評価

さらに詳しい情報をお探しですか?

見積もり、詳細情報をリクエストするか、パンフレットをダウンロードするには、以下のオプションを選択してください。


Features and benefits

pH is one of the main influences on the zeta potential of materials. Understanding how your samples behave over a range of pH conditions is important in determining when they are most, or least, electrostatically stable, or even if their particle size will change (for instance, by agglomeration or dispersion) over a specific pH range. 

The titration method step built into our ZS Xplorer software allows users to quickly and easily perform even complex multi-step titrations, increasing productivity by automating the whole titration process. 

MPT-3 Method NEW.png

Results can be evaluated from within the ZS Xplorer software and any iso-electric points are automatically calculated for you. 

How it works

The MPT-3 Autotitrator incorporates a pump to carry the sample from the reaction vessel to the Zetasizer, pH probe and measurement electronics, and three syringes for three titrants.

The syringes have low volume displacement and well-characterized plunger diameters. This allows the accurate addition of very small amounts of titrant, as low as 0.28 µL, which enables precise targeting of the desired pH value. This is particularly important at pH values close to neutrality.

The Autotitrator supports the use of two concentrations of acid or base in order to generate a very wide pH range. The system will automatically use the appropriate titrant to provide the minimum amount of additive with the maximum accuracy.

The whole measurement sequence can be stored as a method for subsequent re-use, allowing you to standardize your methods for specific materials. 

Use of a degasser for titrants is recommended to prevent air-locks forming in the titrant lines.

仕様

システム構成 Size and zeta potential vs. pH titrations on Zetasizer Advance Series (zeta potential only on Zetasizer Lab model)