| 00:00:00 | The New Malvern Panalytical Zetasizer: Pro and UltraAdaptive Correlation, MADLS and Particle Concentration | 
| 00:01:38 | Overview | 
| 00:03:23 | Zetasizer: Where is it Used? | 
| 00:03:52 | Zetasizer: Where is it Used? | 
| 00:05:12 | What is a Particle? | 
| 00:05:24 | What is a Particle? | 
| 00:05:24 | What is a Particle? | 
| 00:05:37 | What is a Particle? | 
| 00:05:37 | The Zetasizer ZS, Pro and Ultra | 
| 00:05:46 | The Zetasizer ZS, Pro and Ultra | 
| 00:07:00 | A Basic Characterization Workflow | 
| 00:07:20 | A Basic Characterization Workflow | 
| 00:08:51 | A Basic Characterization Workflow | 
| 00:09:22 | A Basic Characterization Workflow | 
| 00:09:54 | A Basic Characterization Workflow | 
| 00:11:21 | Malvern Panalytical Zetasizer | 
| 00:11:29 | Brownian Motion and Particle Si | 
| 00:11:29 | Brownian Motion and Particle Si | 
| 00:11:29 | Components of the DLS | 
| 00:11:29 | The Correlation Function | 
| 00:11:59 | The Correlation Function | 
| 00:12:07 | The Correlation Function | 
| 00:12:25 | The Correlation Function | 
| 00:12:46 | The Correlation Function | 
| 00:12:46 | The Correlation Function | 
| 00:13:01 | The Correlation Function | 
| 00:13:10 | The Correlation Function | 
| 00:13:21 | The Correlation Function | 
| 00:13:34 | The Correlation Function | 
| 00:13:46 | The Correlation Function | 
| 00:14:24 | Sensitive Aggregate Detection | 
| 00:15:41 | Sensitive Aggregate Detection | 
| 00:17:27 | Sensitive Aggregate Detection | 
| 00:18:08 | Thermal Aggregate Detection | 
| 00:19:36 | Sensitive Aggregate Detection | 
| 00:20:11 | Adaptive Correlation | 
| 00:22:20 | Adaptive Correlation | 
| 00:22:20 | Adaptive Correlation | 
| 00:22:20 | Adaptive Correlation | 
| 00:22:28 | Adaptive Correlation | 
| 00:23:03 | Adaptive Correlation | 
| 00:23:20 | Capillary Cell | 
| 00:23:50 | Capillary Cell | 
| 00:24:54 | Sample-Centric Workflow | 
| 00:25:19 | Sample-Centric Workflow | 
| 00:26:33 | What Data Quality Advice Looks Like | 
| 00:27:14 | What Data Quality Advice Looks Like | 
| 00:27:42 | Predictive Parameters | 
| 00:28:10 | Orthogonal Parameters | 
| 00:28:17 | Orthogonal Parameters | 
| 00:28:31 | Orthogonal Parameters | 
| 00:29:40 | Predictive Parameters (KD and A2) | 
| 00:30:35 | Dynamic and Static Light Scattering | 
| 00:30:42 | Dynamic and Static Light Scattering | 
| 00:31:03 | Predictive Parameters (KD and A2) | 
| 00:31:55 | Predictive Parameters (KD and A2) | 
| 00:32:11 | Orthogonal Stability Prediction | 
| 00:32:24 | Zeta Potential | 
| 00:32:30 | Zeta Potential | 
| 00:32:48 | Zeta Potential | 
| 00:33:10 | Zeta Potential | 
| 00:34:29 | Zeta Potential | 
| 00:35:05 | Zeta Potential | 
| 00:35:14 | Zeta Potential | 
| 00:35:25 | Zeta Potential | 
| 00:35:40 | Zeta Potential | 
| 00:36:03 | MADLS® - Multi-Angle Dynamic Light Scattering | 
| 00:36:31 | MADLS® | 
| 00:38:07 | MADLS® | 
| 00:39:03 | MADLS® | 
| 00:39:26 | MADLS® | 
| 00:39:45 | MADLS® | 
| 00:39:54 | MADLS® | 
| 00:41:41 | Particle Concentration | 
| 00:42:29 | Particle Concentration | 
| 00:42:35 | Particle Concentration | 
| 00:42:50 | Allergan UK Adeno-Associated Virus (AAV) Concentration Measurements | 
| 00:43:49 | Allergan UK Adeno-Associated Virus (AAV) Concentration Measurements | 
| 00:44:08 | Allergan UK Adeno-Associated Virus (AAV) Concentration Measurements | 
| 00:44:53 | Allergan UK Adeno-Associated Virus (AAV) Concentration Measurements | 
| 00:45:23 | Allergan UK Adeno-Associated Virus (AAV) Concentration Measurements | 
| 00:45:41 | Allergan UK Adeno-Associated Virus (AAV) Concentration Measurements | 
| 00:46:00 | Particle Concentration | 
| 00:46:52 | Allergan UK Adeno-Associated Virus (AAV) Concentration Measurements | 
| 00:47:19 | Particle Concentration | 
| 00:48:09 | Particle Concentration | 
| 00:48:44 | Particle Concentration | 
| 00:48:52 | The New Zetasizer | 
| 00:52:57 | Thank You for Your Attention | 
| 00:53:05 | Q&A | 
| 01:01:55 | Contact information | 
In this Webinar we will learn about two new exciting additions to a wide range of sample types and concentrations that significantly improve measurement quality, accuracy and sample tolerance through Adaptive Correlation, also enabling sizing now in as little as 24 seconds. Dynamic, Static and Electrophoretic light scattering (DLS, SLS and ELS) are established techniques to accurately measure particle size and charge, amongst many other parameters, over a wide concentration and sample-type range. Malvern Panalytical are global leaders in providing light scattering solutions, demonstrated by the market-leading Zetasizer Nano instrument series. Our new capillary cell now enables users to measure size with as little as 3 μL sample, and with greater accuracy for particles >0.5 μm diameter. Finally, this webinar will describe our unique new measurement type MADLS (Multi-Angle DLS) that improves sizing resolution up to 2:1 and enables the user not only measure particle size, but now also particle concentration in as little as 2 minutes.
Speakers
Patrick King Ph.D. - Field Application Specialist - Bioscience
More information
Who should attend?
Anyone interested in Dynamic Light Scattering techniques such as DLS, SLS and ELS.
Why attend?
To educate yourself on the advances / improvements now possible with the New Zetasizer Ultra or Pro