Too much wear on the plunger tip in the titration syringe can manifest itself as poor ITC data. Best practice is therefore to replace the plunger tip for your MicroCal Auto-iTC200 system at least every 300 experiments or at the first sign of ware. This short video demonstrates how you do it in a few simple steps.