How important is the refractive index of nanoparticles?

Zetasizer Lunch & Learn series - Nano to Advance

Do refractive index and absorption matter for nanoparticles?

Abstract: We discuss the refractive Index of Gold, Silver, Titanium and Other NanoParticles. In this post we also point out refractive index of some polymers and liposomes. Finally, we look at whether the refractive index is needed at all times for DLS.

FAQ: So, How important are refractive index and absorption for nanoparticles?

In general, the optical properties of the scattering material have a tremendous influence on the observed scattering behavior. Mie theory can fully describe these phenomena. Therefore Mie theory is the best choice. For instance, the refractive index n and absorption k of the scattering material affect the scattering intensity. As a result, for many researchers interested in the nanomaterial size range, the optical properties of the material itself are unknown. What can we do in such a case?

Things to consider for optical properties of nanoparticles (Zetasizer)

  1. Firstly, the amount of scattering is directly related to the properties of the material. However, for dynamic light scattering (DLS), the material properties – although often requested in the setup of an experiment – may be irrelevant. But if only an average size by intensity and an average polydispersity (PDI) by intensity are required, then it does not matter which material produced the intensity. The material properties do come into play when the intensity size distribution is transformed into a volume or number distribution. In that case, we have to know exactly how much light is scattered by each nanoparticle. And in order to predict that, Mie theory requires the refractive index and absorption of that particle.
  2. Secondly, for small nanoparticles of less than 100nm, the material properties will not even matter. As a result, the volume distribution obtained via DLS will not change significantly in that case.

Does the refractive index matter for zeta potential?

Likewise for zeta potential, the material properties do not contribute to the calculation of the electrophoretic mobility. Only the dispersant properties come into effect.Therefore in principle, you would not need to enter any parameters for zeta potential measurements. However the software does require selecting a material.

In addition to the above two points, you can model refractive index values and see what happens:

  • For example, you can convince yourself of the effect of the material properties. To clarify how to do this, first edit an existing data record (high-light the record, right-click, edit record). After that, give it a new sample name [for example “Tungsten sample with n=1.6 and abs=0.01”] . Then you can edit the Material properties [click on the dotted box next to it, Add, enter a material name and its associated refractive index and absorption] and then OK OK. As a result, a copy of your original record with the new analysis parameters will appear in your file. By highlighting both records [press the Ctrl-key and highlight both] the results with the two different methods can now be overlayed and subsequently compared.
    You can then confirm that there is no difference in the intensity distribution result (and the z-average and polydispersity for that matter).
    Therefore, you can directly observe the influence that a change in the material properties of the scattering material may have on the volume distribution.
  • In addition, some optical properties for nanomaterials can be found via google. For example, here is a shortlist of refractive index and absorption values for common nanomaterials below. To clarify, this is for a Helium-Neon laser with λ=632nm (the wavelength in the Zetasizer).

In conclusion, in the table below we list a few common materials. Some – however not all – are part of the standard software parameter list. The latest Zetasizer Advance software has a series of standard material properties integrated into the software, and a screen snippet of these values is at the very bottom of this post. [The software snippet contains polystyrene, protein, colloidal gold, SiO2, TiO2, colloidal silver, liposomes, magnetite and hematite.]

Table of select material properties
Sample material  refractive index  absorption
Liposomes #
Phospholipidsn=1.45k=0.001
Exosomesn=1.37 – 1.39*k=0.01
Microvesicles (> .2µm)n=1.40*k=0.01
Nanoparticles and Colloids
Gold [Au]n=0.20k=3.32
Silver [Ag]n=0.135k=3.99
Platinum [Pt]n=2.32k=4.16
Palladium [Pd]n=1.77k=4.29
TiO2n=2.41k=0.001
SiO2n=1.54k=0.00
PFOB emulsionsn=1.305k=0.10
Nanodiamondsn=2.42k=0.00
Macromolecules
Proteinsn=1.45k=0.001
Polystyrenen=1.59k=0.01

# “Optical characterization of liposomes by right-angle light scattering and turbidity” Biochimica et Biophysica Acta (BBA) – Biomembranes 1467, 1,  219-226 (2000)
* “Measurement of refractive index by nanoparticle tracking analysis reveals heterogeneity in extracellular vesicles” Journal of Extracellular Vesicles 2014, 3:25361 DOI: 10.3402/jev.v3.25361 (2014)
+ In addition, the “Refractive index of amorphous polumers” at polymerdatabase has a relevant related list.

In summary, the short answer is: even without the parameters, we can obtain useful information from nanoparticles with DLS.

Finally, if you have any questions, please email me at ulf.nobbmann@malvern.com – Thanks!

PS: a sreen snippet from the latest software material properties is included here for reference. This may be expanded in future versions, and you can add new materials using the pencil icon.

Available refractive index values in zetasizer advance series - these are part of the software installation and always available. Items with the MP logo are supplied by Malvern Panalytical. Other items can be added by yourself with the pencil icon.

If you’re working with Zetasizer Nano and want to improve your understanding of data quality, interpretation, and what comes next, our three-part Lunch & Learn series is designed to guide you through it.👉 Register for the series here >