Zetasizer AT
On-line particle size analyzer for nano-scale processes
The Zetasizer AT brings the industry leading Dynamic Light Scattering (DLS) technology of the Zetasizer Nano to the process development and manufacturing arena, enabling real time monitoring of nanoscale particle processes on-line.
Overview
The Zetasizer AT combines Malvern Panalytical's industry leading Dynamic Light Scattering (DLS) technology with two decades of expertise in design and automation of process interfaces for analytical instrumentation.
The Zetasizer AT is a fully-automated nanoparticle characterization system, delivering lab quality results on-line and in a timely manner for process optimization.
- Ability to monitor processes involving nanoscale particles or droplets in real time e.g. homogenization
- Ability to track changes in particle size with time and determine when the end point is reached
- Automated sampling system facilitates sample extraction, degassing and dilution to the required concentration minimizing operator input
- Sampling system designed to minimize sample consumption
- Easy transfer of laboratory DLS methods to on-line system
- Utilizes Malvern's Universal Interface for system integrators (Malvern Link II)
How it works
An automated sample preparation system utilizes high precision pumps and flow control valves to extract a sample out of a process pipe or vessel and dilute it to the required concentration. The sample is then degassed and pumped through the analyzer’s flow cell. After analysis the sample is discharged and the cycle repeats. Results are immediately available on the screen of the system computer and can be linked to plant SCADA or LIMS system.
Malvern Link II software controls every aspect of the sampling, dilution and degassing steps using Standard Operating Procedures (SOPs).
The Zetasizer AT also allows easy batch measurements of samples from other areas in the process, effectively doubling as a laboratory system.
Specification
- Measurement range:
- 1nm – 1 micron* (diameter) (0.3nm – 10 microns* at-line)
- Measurement principle:
- Dynamic Light Scattering
- Accuracy:
- Better than +/-2% on NIST traceable latex standards
- Precision / Repeatability:
- Better than +/-2% on NIST traceable latex standards
- Sensitivity:
- 0.1mg/mL (Lysozyme)
- Sample Interface:
- 1/16" OD (0.040" ID) or 1/8" OD (1/16" ID) tube
- Pressure:
- 0 - 3 Bar (0- 44 psi) at sample inlet.
- Temperature:
- 0 – 90 °C (flow cell only)
- Minimum sample volume:
- 4ml (application dependent)
- Dilution:
- up to 1:50 / infinite (application dependent)
- Power:
- 110-240VAC, 6A max, 50-60Hz
- Compressed Air for optics purge (optional):
- 2-8 Bar (29-116 psi) per ISO8573-1:2010 [1.7.2]
- Ingress Protection (IP) rating:
- IP54
- Dimensions (W, D, H):
- 98 x 66 x 95 cm (38.6 x 26 x 37.4 inch)
- Weight:
- 120.5 kg (266 lb)
- Conforms to:
- Low Voltage Directive 2006/95/ECEMC Directive 2004/108/ECPressure Equipment Directive 97/23/ECBS EN 61010-1:2010 BS EN 60204-1:2006 + A1:2009 BS EN 61326-1:2006CE marked
Support
Protect your investment and ensure optimized performance at all times with Malvern Panalytical's service plans.
When you purchase a Malvern Panalytical product we understand that this is just the first stage of a working relationship that will last for the lifetime of the instrument. Depending on your needs, Malvern Panalytical will provide the support for your business.
Help Desk
- Telephone and email support.
- Operators who take responsibility for your call and ensure your query is answered.
Training
Software downloads
Malvern Panalytical's service options- choose the right plan for you:
Malvern Panalytical Platinum Plan
For laboratories where maximizing instrument up-time is critical to its daily performance. Our highest level of response and the 'all inclusive' price plan takes care of those unexpected repair bills should the unfortunate occur.
Malvern Panalytical Gold Plan
Enhance your laboratory productivity by maintaining instrument efficiency. Still receive that priority response, combined with specialist technical and software support, we'll keep your instrument performance on track to deliver.
Malvern Panalytical Bronze Visit
For laboratories looking to optimize instrument performance, regular preventative maintenance is essential.
| Platinum Plan | Gold Plan | Bronze Visit | |
|---|---|---|---|
| Annual PM/PV Test | yes | yes | no |
| Telephone / Email Support | yes | yes | no |
| Priority Response | yes | yes | no |
| Emergency Breakdown visits* | yes | yes | no |
| Parts Included | yes | no | no |
| Technical & Software Support | yes | yes | yes |
| IQ/OQ (Pharma)*** | yes | yes | yes |
*including labour & travel costs***available at an additional cost