See how DSC can be used to assess the effects of polyol excipients on antibodies. Differential scanning calorimetry was used to monitor polyol-induced increases in thermal stability across seven monoclonal antibodies. The increases in thermal stability were measured as a function of polyol concentration and orientation of hydroxyl groups. It was found that increases in thermal stability were linear with respect to alcohol concentration. Further, the increase in thermal stability was dependant on the transition temperature; antibodies with lower transition temperatures exhibited a greater increase due to the addition of polyols than antibodies with a higher transition temperature.

Login

Not registered yet?

Sign up for free today. By registering you will have free access to exclusive content including

  • Webinars, presentations and videos
  • Application notes, technical notes, articles, white papers and software downloads

And in addition you will receive

  • Our regular eNews including the latest news, education, events and offers from Malvern Panalytical