The TDAmax consists of three components – the GPCmax integrated solvent and sample delivery module, the TDA 305 Triple Detector Array, and the OmniSEC software. The TDA 305 incorporates RI and, Light Scattering detectors, as well as a viscometer and optional UV detector. Both the integrated detectors and the chromatography columns are fully temperature controlled in an oven to ensure optimum detector stability. In addition, the detectors are connected in series to ensure maximum sensitivity.

  • Absolute molecular weight of small polymers, protein stability and protein aggregation using Right angle light scattering (RALS)
  • Direct output of absolute molecular weight of polymers without extrapolation using Low angle light scattering (LALS)
  • Protein and polymer structure (including branching) information using intrinsic viscosity detector
  • Molecular size in terms of hydrodynamic diameter (Rh) and the radius of gyration (Rg)
  • Copolymer composition using photo-diode array UV detector
  • Autosampler incorporated to increase throughput and improve repeatability
  • Advanced GPC/SEC software capable of going from data to results in just two clicks
  • 3D absorption plot for chemical identification of each component
  • Full suite of calibration and GPC/SEC calculation types

The GPCmax contains an integrated pump, autosampler and degasser. Mobile phase, stored in solvent bottles placed on top of the instrument, enters the dual piston pump after passing through the eluent sensor. This will automatically stop the flow if the solvent is exhausted mid-run. An inline degasser then removes dissolved gases from the mobile phase, which improves the baseline on any of the detectors that may be connected. The pump is compatible with any solvent, even acidic buffers.

The autosampler inside the GPCmax can hold up to 120 vials, which can either be heated, or cooled to prevent degradation of the samples over time. The Autosampler provides the system with variable, reliable, accurate and reproducible injection volumes, enabling comprehensive macromolecular characterizations to be made with confidence.

Mobile phase (and sample) enters the TDA and passes through the GPC/SEC chromatography columns. The columns are maintained at the same temperature as the detectors. After eluting from the column, the dissolved polymer molecules, now separated by size, pass through the UV detector (if present), followed by the light scattering and RI detectors. Finally, the mobile phase passes through the viscometer before going to waste.

System

System:
GPCmax
Pump flow rate:
0.01 to 9.99 mL/min, programmable soft start
Pulsation:
Less than 1%, measured with viscometer
Degasser:
Two channels, < 0.5ppm oxygen at ) 0.5mL/min, eluent sensor to stop pump when eluent runs out
Dimensions (W, D, H):
55cm x 53cm x 38cm
Weight:
37kg
Software:
OmniSEC, 6 channels at 5Hz, unlimited acquisition time

Autosampler

Capacity:
120 vials, programmable sequence
Injection volume:
Variable 20 to 150µL
Reproducibility:
Better than 0.5%
Carryover:
Less than 0.1%, depending on wash program
Cooling option:
4°C to 40°C, limited to 60 vials
Heated tray option:
Up to 65°C limited to 60 vials heated, 60 vials unheated

TDAmax

Detector:
TDAmax
Temperature range:
Ambient to 80°C, all detector cells and columns
Column capacity:
Up to 5 x 30cm columns
Interface:
6 channel digital to OmniSEC software, 2 analogue inputs RI and UV/LS
Dimensions (W, D, H):
41cm x 54cm x 28cm
Weight:
30kg

Detector 1

Detector:
Light scattering
Measurement principle:
RALS 90° angle, LALS 7° angle
Light source:
3mW laser diode, 670nm, temperature controlled
Cell volume:
18µL
Data acquisition rate:
100Hz collection, 5Hz to OmniSEC

Detector 2

Detector:
Refractive index
Measurement principle:
Deflection
Light source:
LED
Cell volume:
12µL
Data acquisition rate:
100Hz collection, 5Hz to OmniSEC

Detector 3

Detector:
Viscometer
Measurement principle:
4 capillary differential Wheatstone bridge configuration
Transducers:
Digital Inert Transducers with over-pressure protection
Cell volume:
18µL
Data acquisition rate:
100Hz collection, 5Hz to OmniSEC

Detector 4 (option)

Detector:
UV, Single wavelength programmable
Wavelength:
190 to 500nm
Light source:
Deuterium
Cell volume:
10µL
Data acquisition rate:
Analog, A-to-D with 5Hz to OmniSEC

Operating environment

Temperature:
15°C – 40°C
Humidity:
35% - 80% non-condensing
Malvern Panalytical - Viscotek accessories include all the components to meet the requirements of your application

Viscotek TDAmax

Viscotek GPCmax

Image

Integrated Pump, Autosampler and Degasser, perfect for size exclusion and gel permeation chromatography.

The GPCmax is a solvent and sample delivery module exclusively for GPC/SEC applications. It provides the reliable and stable chromatographic foundation for the Viscotek advanced detection systems.

More information

Viscotek TDA 305

Image

The standard for GPC/SEC with integrated refractive index viscometer and light scattering detectors.

Research grade temperature controlled GPC/SEC detector module to measure sample concentration, molecular weight and intrinsic viscosity. The TDA is ideal for characterizing synthetic and natural polymers and proteins.

More information

Viscotek UV Detector

Image

Concentration detector, perfect for GPC/SEC, molecular weight determination and compositional analysis.

Accurate measurement of the polymer or protein concentration profile is essential to good molecular weight or structure data. The high-stability Viscotek UV detector is ideally suited for multi-detection GPC/SEC.

More information

OmniSEC software

Image

Powerful, agile, easy to use GPC/SEC software for instrument control, data acquisition, analysis and reporting.

OmniSEC software is an integrated solution for the advanced analysis of proteins and polymers by GPC/SEC, and is at the heart of all Viscotek systems and detectors. Selectable "protein" or "polymer" user interfaces with simplified work flows improve productivity and help you to maximize the information and efficiency of your experiments. Simply load your samples and the software controls all the sequencing, enabling truly automatic unattended operation.

More information

GPC and SEC Columns

Image

GPC/SEC columns for all aqueous and organic applications

A wide range of columns for analytical GPC/SEC of synthetic and natural polymers and proteins.

More information

GPC and SEC Standards

Image

A range of molecular weight standards for different applications of GPC/SEC

Malvern Panalytical offers a wide range of molecular weight standards for GPC/SEC for conventional and universal calibration as well as light scattering.

More information

Protect your investment and ensure optimized performance at all times with Malvern Panalytical's service plans.

When you purchase a Malvern Panalytical product we understand that this is just the first stage of a working relationship that will last for the lifetime of the instrument. Depending on your needs, Malvern Panalytical will provide the support for your business.

Help Desk

Training

Software downloads

Malvern Panalytical's service options- choose the right plan for you:

Malvern Panalytical Platinum Plan

For laboratories where maximizing instrument up-time is critical to its daily performance. Our highest level of response and the 'all inclusive' price plan takes care of those unexpected repair bills should the unfortunate occur.

Malvern Panalytical Gold Plan

Enhance your laboratory productivity by maintaining instrument efficiency. Still receive that priority response, combined with specialist technical and software support, we'll keep your instrument performance on track to deliver.

Malvern Panalytical Bronze Visit

For laboratories looking to optimize instrument performance, regular preventative maintenance is essential.

Platinum PlanGold PlanBronze Visit
Annual PM/PV Test
yes
yes
no
Telephone / Email Support
yes
yes
no
Priority Response
yes
yes
no
Emergency Breakdown visits*
yes
yes
no
Parts Included
yes
no
no
Technical & Software Support
yes
yes
yes
IQ/OQ (Pharma)***
yes
yes
yes

*including labour & travel costs***available at an additional cost